Broad frequency range 3.10-5 .. 2.107 Hz including low frequencies for characterization of dielectric relaxations, electrochemical and high impedance effects.

Ultra wide impedance range 10-3 .. 1015 Ω (18 decades) covers range from conductors to best isolators.

Ultra wide capacity range 10-15 .. 1 F allows broadband measurement of smallest capacities down to 1 fF.

High phase 2.10-3 ° and loss factor tan(δ) absolute accuracy 3.10-5 for low loss dielectric materials and isolators broadband characterization.

Supports a series of test interfaces for customized applications with special functionality.

Supports non linear Dielectric, Conductivity and Impedance Spectroscopy or gain phase measurements in combination with all test interfaces.

Supports high speed measurements rates up to 157 impedance or 210 gain phase measurement points per second via the GPIB port for online monitoring of time variant processes in the 5 ms range by mainframe option F.

New test interface concept with adaptive system functionality and single port operation by high level GPIB commands for easy and straightforward implementation in own programs or existing measurement set-ups.

Automatic self calibration and diagnosis by user cancels out long term internal drift and verifies functionality.

Precision digital frequency response analyzer up to 40 MHz for two channel gain phase measurements with 0.001° phase - and 10-5 in amplitude resolution included. Operates like a broadband lockin amplifier with two channels, extended accuracy and frequency range.

Optional control WinDETA software for turnkey calibration, operation, data evaluation, 2- and 3- dimensional graphical representation including non linear spectroscopy.